Catalog Number 293001-001
Device Problems Premature Discharge of Battery (1057); Device-Device Incompatibility (2919)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the external battery lost its charge after 20 minutes, the companion 2 driver continued to perform its life-sustaining functions.Also, the companion 2 driver has a redundant, alternate power source of external wall power.The companion external battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion external battery could only be used in the companion 2 driver for 20 minutes before losing its charge.There was no reported adverse patient impact.
 
Manufacturer Narrative
Visual inspection and system management (smbus) data review revealed no damage or abnormalities.The companion external battery passed all functional testing requirements including the required 30 minute discharge test.The battery performed as intended with no evidence of a device malfunction.Therefore, the root cause of the customer-reported early discharge could not be determined.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.Ce 4340 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION EXTERNAL LI-BATTERY PACK
Type of Device
BATTERY PACK
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7586805
MDR Text Key111146990
Report Number3003761017-2018-00226
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Remedial Action Replace
Type of Report Initial,Followup
Report Date 04/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number293001-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2018
Was the Report Sent to FDA? No
Date Manufacturer Received05/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age61 YR
-
-