• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37602
Device Problems Electromagnetic Interference (1194); Failure to Deliver Energy (1211); Low Battery (2584); Device Displays Incorrect Message (2591); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/28/2016
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient who was implanted with an implantable neuro stimulator (ins) for essential tremor.The patient reported an elective replacement indicator (eri) and wanted to know what the eri meant.Patient stated that they did not know what they got near but it shut off, patient did not remember the date but thinks it was because they had to go to a police station and walked through one of those things, further stating that they think that had to be what shut the device off because it was out of the ordinary when it happened.Patient stated it was almost two years ago after that was when it shut off.Patient was also dissatisfied with longevity of the ins , further mentioning that before their implants always lasted like four years and this one did not even last two years.Patient saw an eri on (b)(46 and the device was replaced in (b)(6) 2016.Patient was also dissatisfied with longevity of the ins , further mentioning that before, their implants always lasted like four years and this one did not even last two years.No patient symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7587071
MDR Text Key110621812
Report Number3004209178-2018-13073
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00613994761057
UDI-Public00613994761057
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2015
Device Model Number37602
Device Catalogue Number37602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/07/2018
Date Device Manufactured07/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
-
-