Model Number EGIA60AMT
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic right hemicolectomy procedure, at resection of ascending colon cancer, at functional anastomosis.After the firings, the surgeon found a yellow broken piece in the cavity by the camera and was able to retrieve it.No patient injury was reported.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device with articulating medium/thick reload.Visual inspection of the returned product noted that the reload was fully fired.Functional evaluation of the reload found no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mc dermott rd
north haven CT 06473
MDR Report Key7588120
MDR Text Key110664852
Report Number1219930-2018-03009
Device Sequence Number1
Product Code GDW
UDI-Device Identifier30884523003219
UDI-Public30884523003219
Combination Product (y/n)N
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberEGIA60AMT
Device Catalogue NumberEGIA60AMT
Device Lot NumberN8A0096KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2018
Date Manufacturer Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-