Catalog Number BI70002000-G30
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Unique device identifier (udi) is unavailable.Device manufacture date is unavailable.A medtronic representative went to the site to test the equipment.Testing revealed that the 400v were not present on the outbound side of the power enclosure.Replacing the enclosure resolved the issue.The imaging system then passed the system checkout and was found to be fully functional.The power enclosure has been received by the manufacturer for evaluation however results are not available yet.
 
Event Description
Medtronic received information regarding an imaging device being used outside of procedure.It was reported that after the imaging system was installing, the system was stuck in the initializing phase.There was a loss of 400 v out of the power conversion enclosure.There was no patient involvement.
 
Manufacturer Narrative
The analysis on the power conversion enclosure was completed by medtronic personnel.Testing found that a transistor on the circuit board of the unit was shorted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7588453
MDR Text Key110709446
Report Number1723170-2018-02676
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000-G30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received09/05/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-