• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926016250
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the stent delivery system (sds) was returned for analysis.A visual examination of the stent found no issues on the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter (od) was measured and is within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.The tip was visually and microscopically examined and no signs of damage were noted.A visual and tactile examination found a hypotube break at 650 mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted during device examination.This type of damage most likely occurred due to excessive forces that could have been applied on the delivery system.A visual and tactile examination of the outer and mid-shaft found no issues with the extrusion shaft.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event Description
Reportable based on analysis completed on 18-may-2017.It was reported that shaft kink occurred.A 2.50 x 16mm synergy ii drug-eluting stent was advanced to treat the lesion.However, when the device was put into the guide, the shaft was kinked.The procedure was completed with another synergy stent.No patient complications were reported and the patient's status was fine.However, returned device analysis revealed hypotube break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7589963
MDR Text Key110707073
Report Number2134265-2018-05000
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729840091
UDI-Public08714729840091
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH7493926016250
Device Catalogue Number39260-1625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-