The device was returned for analysis.The engineering evaluation result stated that no knob nor connected deployment line was returned.At approximately 3 cm from hub, there was columnar failure of dual lumen.There was 1 cm of dual lumen stretched out near the transition.Approximately 5 cm of distal shaft, upon which the endoprosthesis is mounted, was exposed at transition.The endoprosthesis was longitudinally compressed and displaced on shaft towards the tip.The outer braided constraining line was completely deployed with approximately 1 cm of the inner braided constraining line deployed.The endoprosthesis was partially expanded with approximately 5.7 cm of endoprosthesis still constrained.The deployment line connected to the endoprosthesis went into the transition port.There was a broken fiber in the deployment line approximately 12 cm from endoprosthesis.Deployment was able to be continued with traction of the deployment line at the endoprosthesis.Based on the device examination performed, no manufacturing anomalies were identified.Per the gore® viabahn® endoprosthesis instructions for use (ifu), it is stated that complications and adverse events that can occur when using any endovascular device.These complications include but are not limited to deployment failure.A warning is given as ¿inadvertent, partial, or failed deployment of the endoprosthesis may require surgical intervention.¿.
|