Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the pin driver was broken while pulling the pins out, and the broken piece was retrieved.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
Udi # - (b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned part indicates that the device exhibits signs of repeated use and the tip of the driver has fractured off.Dimensional examination indicates that the driver conforms to the measurable print specifications.Sem analysis indicates that the driver fractured due to overload.Quasi-cleavage mode of fracture identified throughout the fracture surface.Fracture surface revealed multiple crack exit sites near the outer diameter of the pin driver and that indicates the cracks propagated towards its outer diameter after initiating near the inner diameter of the driver.Dhr was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL PIN AND SCREW INSERTER
Type of Device
EXTRACTOR
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7590463
MDR Text Key110703677
Report Number0001822565-2018-03072
Device Sequence Number1
Product Code HWB
Combination Product (y/n)N
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00590102100
Device Lot Number63373630
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Was the Report Sent to FDA? No
Date Manufacturer Received02/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age72 YR
Patient Weight78
-
-