Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Information (3190)
|
Event Date 05/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the pin driver was broken while pulling the pins out, and the broken piece was retrieved.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
Udi # - (b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned part indicates that the device exhibits signs of repeated use and the tip of the driver has fractured off.Dimensional examination indicates that the driver conforms to the measurable print specifications.Sem analysis indicates that the driver fractured due to overload.Quasi-cleavage mode of fracture identified throughout the fracture surface.Fracture surface revealed multiple crack exit sites near the outer diameter of the pin driver and that indicates the cracks propagated towards its outer diameter after initiating near the inner diameter of the driver.Dhr was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|