(b)(4).Batch # p94l10.Device analysis: the analysis results found that the 2b5lt device was returned with the obturator cannula damaged bent and inserted through the sleeve.In addition, the tyvek (p4t43r) from another device was returned along with the instrument.Upon testing, the obturator was removed with difficulties through the sleeve.The device was disassembled, in order to evaluate the condition of the seals and was confirmed that the seals were found damage and the lubricant was insufficient from the seals causing the removing issue.The most likely cause for the found condition is related to the manufacturing process.The reported complaint was confirmed.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|