(b)(4).A field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse found the hemolysis & wash (h/w) tubing sinker was not submerged in the h/w and dilution well.The fse also found a lack of wash entering the system.The fse cleaned the dilution well and ran samples to test operation.The fse ran 20 samples without error; calibrated the analyzer, then ran quality controls and precision.All results were within acceptable range.All errors cleared and the g8 instrument was operational.No further actions were required by field service.A 13-month complaint / service history review for similar complaints was performed for the serial (b)(4) from aware date of (b)(6) 2018, which included data from 4/21/2017 through (b)(6) 2018.There were two similar complaints identified during the searched period for low total areas and there were two similar complaints identified during the searched period for missing peaks, which includes this event.The g8 operator's manual under chapter 1 and chapter 6 troubleshooting are as follows: 1.4 storage and stability columns are warranted for 2500 injections.1.8 limitations of the procedure total area dilution studies demonstrate that the assay is linear from a total area of 500 to 4000.However, the optimum total area is 700 to 3000.The 200 area low error three successive results below the lower limit of the total area (50) occur.If the error message is present when sufficient volume of sample is set in the rack, the problem may be caused by an empty reagent (hemolysis and wash solution).Check the remaining volume of hemolysis and wash solution and start the assay again.Error message: 200 area low error explanation: the peak area which does not reach the minimum required area (50) occurred three times in series.Countermeasure: check the samples, buffers, and hemolysis and wash solution.The 220 no peak detect no peaks could be detected.This problem could be caused by different scenarios.Insufficient sample uptake due to a coagulated sample may have been processed or a short sample was processed.The most probable cause for the reported event was the column count (cc) was greater than 2500.The cc was 2840.The h/w tubing was not submerged in the wash.
|