(b)(4).A field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse verified the reported event and completed a preventive maintenance.The issue was resolved.The fse found a clog in the injection valve area.The fse verified precision, ran calibration, and ran quality control.All results were within acceptable range.All errors cleared and the g8 instrument was operational.No further actions were required by field service.A 13-month complaint / service history (b)(4) review for similar complaints was performed for the serial number (b)(4) from aware date of 5/19/2018, which included data from 4/19/2017 through 5/19/2018.There were no similar complaints identified during the searched period, which includes this event.The g8 operator's manual codes under chapter 6 troubleshooting are as follows: 100 pressure high - the pump pressure exceeded the upper limit (15 mpa) set in the pres high parameters.When the filter or column replacement period has been exceeded, first replace the filter or column.If the pressure is still high, remove the inlet and outlet flow line around the column and filter, and determine which part is the cause of the high pressure.Then, contact a technical support representative.If the pressure displayed on the screen is: (a) greater than the pressure on the column inspection report + 4 mpa, then replace the filter.(b) less than the pressure on the column inspection report, then proceed with priming the column; 102 pres limit over - the pump pressure has risen abnormally and has caused the shutdown circuit to be activated.Turn the main power switch off and remove the cause of the pressure increase.Refer to 100 pressure high.The most probable cause for the reported event was a clog in the injection valve.
|