Qn#(b)(4).The customer returned one unit ae05ml autoend05 ml for investigation.The sample is for this complaint tc and tc 1900060830.The returned sample was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appears used as there is biological material present on the device.Reference file (b)(4) for investigation photos.Functional inspection was performed on the returned sample by attempting to engage the trigger using hand pressure.Upon engagement of the trigger, an audible ratchet sound was heard indicating that the internal ratchet ears are intact.The first clip was able to properly load into the jaws of the applier and was successfully applied to over-stressed surgical tubing.This was repeated with the same result.The sample was received with 2 clips remaining in the channel indicating that 13 clips were fired by the end user.No defects or anomalies were observed.The ifu for this product, l06072, was reviewed as a part of this complaint investigation.The ifu for this product states, "mishandling of appliers may result in improper load and/or closure of the ligating clip." other remarks: the reported complaint of "jaws not closing" was not confirmed based upon the sample received.One device was returned.Upon functional inspection, no problems were found as clips could be loaded into the jaws and close with no issues.The jaws were able to open and close properly.The device was received with 2 clips remaining in the channel indicating that the end user fired 13 clips.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.No functional issues were found with the returned device.Additional method code: 26.
|