• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEW WORLD MEDICAL, INC. AHMED GLAUCOMA VALVE; GLAUCOMA SHUNT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEW WORLD MEDICAL, INC. AHMED GLAUCOMA VALVE; GLAUCOMA SHUNT Back to Search Results
Model Number FP7
Device Problem Failure to Prime (1492)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
The device history record for the lot reported was reviewed and no issues were observed the product was manufactured, tested and released in compliance with procedures.Customer was contacted and asked for the sample for evaluation without a sample for evaluation we are unable to determine what caused the issue reported.
 
Event Description
Device failed to irrigate properly.No injury to patient.Device removed immediately from the field and replaced.
 
Manufacturer Narrative
Customer provided unused valve for evaluation.The unit was visually and functionally tested and it performed in compliance with specification.The unit was first primed as indicated in the ifu and no issues were observed.A steady state test was completed.The valve was able to maintain the required pressure over time.
 
Event Description
Device failed to irrigate properly.No injury to patient.Device removed immediately from the field and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHMED GLAUCOMA VALVE
Type of Device
GLAUCOMA SHUNT
Manufacturer (Section D)
NEW WORLD MEDICAL, INC.
10763 edison court
rancho cucamonga CA 91730
Manufacturer (Section G)
NEW WORLD MEDICAL, INC.
10763 edison court
rancho cucamonga CA 91730
Manufacturer Contact
cristina avalos
10763 edison court
rancho cucamonga, CA 91730
9094664304
MDR Report Key7591867
MDR Text Key111307477
Report Number1000125279-2018-00011
Device Sequence Number1
Product Code KYF
UDI-Device Identifier00892064002119
UDI-Public(01)00892064002119(11)180409(17)200409(10)D0218
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162060
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/09/2020
Device Model NumberFP7
Device Catalogue NumberFP7
Device Lot NumberD0218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-