• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; VISCOMETER, MUCUS, CERVICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; VISCOMETER, MUCUS, CERVICAL Back to Search Results
Model Number 383400
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Date 05/16/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reported the enteral feeding pump failed the rotor error alarm test.The unit did not alarm at all.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed and the customer states, ¿pump failed the rotor error alarm test.¿ the unit was triaged and the customer¿s reported condition was confirmed.A trend has been identified and a corrective and preventative action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
VISCOMETER, MUCUS, CERVICAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7592080
MDR Text Key110770001
Report Number1282497-2018-00479
Device Sequence Number1
Product Code LHZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/21/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-