It was reported that balloon pinhole occurred.A 3.50mm x 8mm nc emerge® balloon catheter was advanced for dilatation.However, upon inflation, it was noted that the contrast media leaked into the artery and a hole was noted on the balloon.The procedure was completed with another nc emerge balloon catheter.There were no patient complications.
|
Device evaluated by mfr.: returned product consisted of a nc emerge balloon catheter.The balloon was loose and there was fluid in the balloon and inflation lumen.The balloon, hypotube, inner and outer shaft were microscopically inspected.Inspection revealed tip damage, abrasions in the balloon material located over the distal markerband, and a perforation in the balloon material approximately 1mm in length (also located at the distal end of the markerband).The abrasions are consistent with difficult anatomy.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.There is no indication the device was inflated over rated burst pressure (rbp).The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
It was reported that balloon pinhole occurred.A 3.50mm x 8mm nc emerge balloon catheter was advanced for dilatation.However, upon inflation, it was noted that the contrast media leaked into the artery and a hole was noted on the balloon.The procedure was completed with another nc emerge balloon catheter.There were no patient complications.
|