TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: field service engineer (fse) visited the customer to address the reported event.During evaluation, the fse was able to confirm the z1-axis errors on print out.Fse checked all alignments to sample.Fse lubricated z and y axis with triflow.Fse installed a new sample needle and checked to make sure needle was not hitting inside of lucite block.Fse ran controls, precision, and 2 racks of patients.The instrument was verified as operational.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were no similar complaints found during this time.The g8 operator's manual under chapter 5 maintenance procedures, 5.10 sampling needle replacement has a caution that states the following: replace the sampling needle if it is bent or broken.Use the following procedure to replace the sampling needle.Access to the inside of the analyzer is needed to replace the sampling needle.Be sure that only personnel who have been trained by tosoh or its representatives perform these operations.Be sure to wear protective clothing (goggles, gloves, mask, etc.) and take sufficient care to prevent infection during handling.Take care not to touch the end of the sampling needle during handling.If the needle placement is clearly off center of the primary tube, it must be adjusted.Cancel the assay and contact technical support.The most probable cause of the bent sample needle was due to lack of lubrication of the y-axis.
 
Event Description
A customer reported that the sample probe bent when it hit the lucite block while using on the g8 instrument.The customer changed the sample probe and the issue reoccurred.The second probe also hit the lucite block and bent.Technical support specialists sent the customer a replacement sample probe to install.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7594961
MDR Text Key111444739
Report Number8031673-2018-00534
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/04/2018
Distributor Facility Aware Date05/24/2018
Device Age3 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer06/04/2018
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-