Model Number N/A |
Device Problem
Difficult To Position (1467)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/07/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a right hip procedure, the surgeon impacted the liner into the cup and the liner did not seat.After several attempts, the surgeon requested a second liner to complete the surgery.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned liner shows damage from attempted implantation and removal confirming the complaint.The returned device also shows scallops damage and deep gouges and scratches on the inner lip diameter.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|