(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
It was reported from (b)(6) that during service and evaluation, it was determined that the motor of the small battery drive device seized, jammed, and was moving heavily.It was noted that the motor was moving heavily and the electronic control unit (ecu) magnetic plate detached out.It was further determined that the device failed pretest for check the triggers and ecu function, check switching function, check the oscillation angle, and check the off/oscillation/on switch mode function.It was noted in the service order that the product could not be used.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|