MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problems Device Inoperable (1663); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative confirmed the issue and reported that the dongle would not attach to the panel and found that the standoffs were damaged.The broken screw shafts from mounting cable were removed.The representative reported that standoffs were replaced and dongle was reattached to resolve the issue.The rear cab breakout panel was returned to the manufacturer for evaluation.Through visual and physical testing it was confirmed that two standoffs screws were missing from the replacement.The standoff screws were removed from the replacement rear cab breakout panel and put into the panel that had the broken standoff screws in order to fix the problem.
 
Event Description
Medtronic received information regarding an imaging system.It was reported that the imaging system was booting but was stuck on e-stop.While troubleshooting it was found that the dongle was disconnected.When the dongle was reset, the e-stop was able to be cleared however, the dongle appeared to have been bumped and would not stay seated in the screws.No patient was present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7599024
MDR Text Key111005533
Report Number1723170-2018-02768
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received11/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-