Detachment Of Device Component (1104); Premature Activation (1484); Physical Resistance (2578)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.The device was implanted into the patient.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure in the splenic artery using pod packing coils (podjs).During the procedure, the physician successfully placed multiple coils using a non-penumbra microcatheter.The physician then felt resistance while advancing a podj and therefore attempted to retract the podj.While retracting, the podj unintentionally detached within the microcatheter.The physician was able to push the podj out of the microcatheter and into the patient, and the procedure ended at this point because the existing embolization was sufficient.There was no report of an adverse effect to the patient.
|
|