• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. POD PACKING COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD PACKING COIL; HCG, KRD Back to Search Results
Device Problems Detachment Of Device Component (1104); Premature Activation (1484); Physical Resistance (2578)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/14/2018
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.The device was implanted into the patient.
 
Event Description
The patient was undergoing a coil embolization procedure in the splenic artery using pod packing coils (podjs).During the procedure, the physician successfully placed multiple coils using a non-penumbra microcatheter.The physician then felt resistance while advancing a podj and therefore attempted to retract the podj.While retracting, the podj unintentionally detached within the microcatheter.The physician was able to push the podj out of the microcatheter and into the patient, and the procedure ended at this point because the existing embolization was sufficient.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7601043
MDR Text Key111043956
Report Number3005168196-2018-01181
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
-
-