(b)(4) the complaint device was not returned to fisher & paykel healthcare (b)(4) for evaluation.The hospital had informed us that the heated breathing tube (hbt) used was an 900pt551 airspiral hbt.Following this complaint we approached the hospital for further information, it was reported that no damage to the cannula or the circuit was observed.We were further informed that the patient: had to be bagged and masked with 100% oxygen to maintain saturation and later intubated.Was very fidgety and may have disconnected interface himself.Was very sick and they were expecting to intubate him - disconnection from airvo 2 may have just resulted in intubation happening sooner.Conclusion: the customer reported no damage to the cannula or the circuit.Thus, this suggest that there was no fault with the complaint device.It is possible that the cannula or the heated breathing tube were accidentally pulled or the patient was trying to remove them.The 900pt551 hbt forms a tight fit with the opt800 and opt900 series cannulae and requires over 15 newtons of force in order to disconnect it from the cannula.The user instructions (ui) for the airvo 2 system specifically warn that the airvo2 "is not intended for life support" and that "appropriate patient monitoring must be used at all times".The ui also instructs the caregiver to "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply." during manufacture, samples of the hbt connector are tested with the opt series cannulae to ensure that the required disconnection force is greater than 15 newtons.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and stretching or deformation.Any product that fails the visual inspection is disposed of.The setup instructions in the user instructions which accompany the opt944 nasal cannulae also warn that "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." our f&p representative has arranged to provide further training to the hospital in setup and fitting of the cannula.
|