• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4469C
Device Problem Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Event date was sometime in (b)(6) 2018.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that two (2) homechoice cassettes were found ¿unglued¿.This issue occurred during use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Two (2) actual samples were received for evaluation and an evaluation is complete.A visual inspection was performed with issues noted.A defective weld was verified on each sample.A batch review was conducted and there were no deviations found related to the reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was determined to be manufacturing related.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7606800
MDR Text Key111879469
Report Number1416980-2018-03642
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K923065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4469C
Device Lot NumberH17J13020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/20/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-