Model Number 8637-40
Device Problem Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Interrogation of the pump determined it was used to infuse remodulin 10mg/ml at 0.062mg/day.The pump was returned, and analysis found a hole in the pump tube/mechanical wear of the pumphead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was provided by a healthcare professional (hcp) via a manufacturer¿s representative regarding an implantable pump.The patient was receiving remodulin 10mg/ml for a total dose of 0.062mg/day.The pump was returned to the manufacturer without a complaint.The reason for pump explant was noted to be due to elective replacement indicator (eri).No patient symptoms or complications were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7609384
MDR Text Key111342315
Report Number3004209178-2018-13606
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/14/2013
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2018
Date Manufacturer Received06/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No