Model Number M0068318170
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2018
Event Type  Injury  
Manufacturer Narrative
The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a sacrospinous ligament fixation procedure on (b)(6) 2018.According to the complainant, during the procedure, the capio cage would not catch the dart.The device was deployed several times then the capio carrier broke.The procedure was completed with another uphold lite with capio slim device.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual examination of the returned uphold lite with capio slim revealed that the capio slim head was separated at the proximal end most likely caused by torqueing the device during use.All rivets were attached.During analysis, the capio head was opened to examine the carrier.There was heavy debris inside the capio head.The carrier was not broken.Analysis revealed no damage to the uphold lite mesh assembly.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu)/product label.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a sacrospinous ligament fixation procedure on (b)(6) 2018.According to the complainant, during the procedure, the capio cage would not catch the dart.The device was deployed several times then the capio carrier broke.The procedure was completed with another uphold lite with capio slim device.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
MDR Report Key7609601
MDR Text Key111338248
Report Number3005099803-2018-01817
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/26/2020
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000060409
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Date Manufacturer Received07/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age78 YR
Patient Weight60
-
-