MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37612
Device Problems Failure to Deliver Energy (1211); Energy Output Problem (1431); Device Displays Incorrect Message (2591); Battery Problem (2885)
Patient Problems Therapeutic Effects, Unexpected (2099); Shaking/Tremors (2515)
Event Date 06/13/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient with an implanted neurostimulator (ins) for essential tremor and movement disorders.The caller reported that the patient's implant was off and their hands were shaking and all over the place when they woke up this morning.The caller reported syncing the ins with the patient programmer which showed the ins needed to be charged.The caller reported they were able to charge the implant, with 8 bars.The caller reported that the recharger was showing less than 1/4 charge.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Information references the main component of the system and other applicable components.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the patient indicating that the device does not turn off overnight, no matter how many hours patient sits to program it.The device does not stop the tremor totally, patient further stated that sometime the device goes off and they sit for a long period of time programming it.Patient stated that they called the manufacturer who walked them through turning it on and patient could not do it on their own since the patient's hand has never totally stopped the tremor, so the wife does it for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7611539
MDR Text Key111405592
Report Number3004209178-2018-13700
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169529762
UDI-Public00643169529762
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2016
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/30/2018
Date Device Manufactured01/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
Patient Weight73
-
-