Catalog Number BI70002000
Device Problems Device Inoperable (1663); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/22/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the mvs computer was taking an inordinate amount of time to boot into the application.The computer would remain in the post screen for approximately 20 minutes before it would eventually proceed.During troubleshooting, representative discovered that the power connection for the external dvd rom drive had become partially disconnected.After re-seating the power connection for the external drive, and the subsequent boot attempts of the mvs computer produced expected results.The imaging system then passed the system checkout and was found to be fully functional.The suspected computer was returned as unused.
 
Event Description
Medtronic received information regarding an imaging device being used outside of a procedure.It was reported that the mobile view station (mvs) computer would not boot into the os.This was discovered during a planned maintenance (pm).No patient was present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
koko tyler
navigation customer quality
826 coal creek circle
louisville, CO 80027
7208903200
MDR Report Key7611876
MDR Text Key111434412
Report Number1723170-2018-02843
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse