MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL; SEE H10) Back to Search Results
Model Number S7
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.No procode, common device name and/or 510k provided as this device is not released for distribution in the united states.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used for a sacroiliac and thoracolumbar fusion procedure.It was reported that preoperatively with no patient present, the site attempted to start the navigation system but failed as the system indicated ¿no signal.¿ several restarts of the system were attempted.The navigation system eventually started but worked in an unstable manner and was noted to be unresponsive.Thus, use of the system was discontinued prior to being used in the procedure.Instead the procedure was completed with the use of a c-arm.There was no surgical delay due to the reported issue, and there was no known impact on patient outcome.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the computer of the navigation system was replaced to restore functionality.The system then passed the system checkout and was found to be fully functional.The computer for the navigation system was returned to the manufacturer for analysis.The computer was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CART 9734056 S7 STAFF SHRT 100-120V INTL
Type of Device
SEE H10)
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7614863
MDR Text Key111988652
Report Number1723170-2018-02863
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10)
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9734056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-