BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER PRIME¿ XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004EPTP4500THN40 |
Device Problem
Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: the device was returned for analysis.Visual inspection revealed that the device has a kink at the distal section approximately at 1.2 cm from the distal tip while in the neutral position and broken adhesive was observed in ring #1.Functional inspection revealed that the steering knob and the tension control knob functioned properly on both lock and unlock positions and no abnormal resistance was felt when actuating the steering mechanism.X-ray inspection revealed a bent center support under the same location of the kink at the distal section.A steering wire partially detached from the center support due to broken solder joint was also observed.Distal end was dissected and bent center support was confirmed between ring 1 and ring 2 and a damage in the kevlar wrap.The distal tip was dissected, the steering wire was partially separated from the center support due to a broken solder joint.Dried body fluid was found within the distal end and the kevlar wrap was found retracted.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is design specification as the device problem was traced back to the design specifications.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on 07-june-2018.It was reported that the catheter tip were kinked.A 7-110-2.5-8-8 n4 blazer prime¿ xp ablation catheter was used for a typical atrial flutter ablation procedure.A non-bsc sheath was used to pass and support the catheter in the low right atrium region to perform a cavo-tricuspid isthmus (cti) line.Then, the catheter tip was noted to be kinked between the second and third electrode after being used for short period in the cti.The catheter was withdrawn and the procedure was completed with a different device.No patient complications were reported.However, returned device analysis revealed a body fluid ingress within the distal end.
|
|
Search Alerts/Recalls
|
|
|