• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120 HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE AUTOSAMPLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120 HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE AUTOSAMPLER Back to Search Results
Model Number ADVIA 2120 HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE AUTOSAMPLER
Device Problem Device Handling Problem (3265)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
The customer contacted siemens customer care center (ccc).Siemens determined that the cause of the needlestick injury to the customer's finger on the advia 2120 hematology system with single aspirate autosampler was the customer's technique when performing the centering collar maintenance on the system.The advia 120/ 2120/ 2120i hematology system operator's guide states the following: "to avoid personal injury and exposure to a potential biohazard, you must cover the needle with the red needle cover immediately after you remove the centering collar." the customer failed to follow the operator's guide instructions and did not cover the autosampler needle with the red needle cover.The system is performing according to specifications.No further evaluation of this device is required.
 
Event Description
The customer sustained a needlestick injury to the finger when performing maintenance on the centering collar on an advia 2120 hematology system with single aspirate autosampler.The customer disinfected the affected area, received first aid treatment, and will be tested for infectious diseases ((b)(6), (b)(6), alanine transaminase (alt) and aspartate transaminase (ast)).There are no known reports of adverse health consequences due to the customer sustaining a needlestick injury while performing maintenance on the centering collar on the advia 2120 hematology system with single aspirate autosampler.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 2120 HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE AUTOSAMPLER
Type of Device
ADVIA 2120 HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE AUTOSAMPLER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
chapel lane, swords, co.
registration number: 8020888
dublin,
EI  
Manufacturer Contact
christina lam
511 benedict ave
tarrytown, NY 10591
9145243504
MDR Report Key7616396
MDR Text Key111678785
Report Number2432235-2018-00260
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K162977
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA 2120 HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE AUTOSAMPLER
Device Catalogue Number10316162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/03/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-