• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 ACETABULAR LINER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 ACETABULAR LINER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during surgery, the liner would not seat into the shell.The surgery was finished with a second liner.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Upon visual inspection the liner has been fractured.However, no further analysis was performed due to the impaction to the liner.There is circular gouging on the liner.There is no damage to the outside of the shell however the inside diameter has deep gouging and damages to the locking feature.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during surgery, the liner would not seat into the shell.The shell was removed and the surgery was finished with a second shell and liner.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 ACETABULAR LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7616572
MDR Text Key112169325
Report Number0001825034-2018-04137
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/06/2023
Device Model NumberN/A
Device Catalogue Number010000848
Device Lot Number6207321
Other Device ID Number(01) 0 0880304 52634 1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Date Manufacturer Received09/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
110010243 6234846 G7 OSSEOTI 3 HOLE SHELL 50MM D; UNKNOWN HEAD; UNKNOWN STEM
Patient Outcome(s) Required Intervention;
-
-