Model Number N/A |
Device Problem
Difficult To Position (1467)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during surgery, the liner would not seat into the shell.The surgery was finished with a second liner.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Upon visual inspection the liner has been fractured.However, no further analysis was performed due to the impaction to the liner.There is circular gouging on the liner.There is no damage to the outside of the shell however the inside diameter has deep gouging and damages to the locking feature.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that during surgery, the liner would not seat into the shell.The shell was removed and the surgery was finished with a second shell and liner.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
|
|