Boston scientific received information that at the device check one week post-implant, this right ventricular (rv) lead exhibited increased pacing threshold measurements and a decrease in pacing impedance measurements.An x-ray showed the lead had moved slightly.A revision procedure was planned to reposition the lead the following day.At the revision, the helix could not be fully retracted.During manipulation, the lead was released from the heart tissue and was removed from the patient.Visual inspection noted tissue caught in the helix; the tissue was removed as much as possible, but the helix still did not function properly, requiring 20 turns to extend when the lead was straight.Therefore, a new lead of the same model was implanted without complication.No additional adverse patient effects were reported.
|
Upon receipt at our post market quality assurance laboratory, visual inspection of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing did not identify any lead characteristics that would have resulted in the clinical observation of dislodgement.
|