Boston scientific received information that, in (b)(6) 2018, this patient presented for a routine generator change-out of this pacemaker.During the revision, a competitor wrench was used to loosen the set-screws of the device.One of the set-screws would not loosen and two wrenches were broken during attempts.The physician made the decision to abandon the system on the left side and implant a new system on the right.A few months later, a second procedure was performed to remove the abandoned system.The physician noted during this procedure that a significant amount of blood was in the device header.The competitors right ventricular lead was cut in order to remove the device.This pacemaker was explanted.No additional adverse patient effects were reported.
|