BOSTON SCIENTIFIC - MARLBOROUGH SYMPHION¿ TISSUE REMOVAL SYSTEM RESECTING DEVICE; INSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA Back to Search Results
Model Number FG-0201
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Uterine Perforation (2121)
Event Date 05/31/2018
Event Type  Injury  
Manufacturer Narrative
The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2018 that a symphion resecting device used in a hysteroscopy procedure performed on (b)(6) 2018.According to the complainant, during procedure and inside the patient, the resecting device was used to do an endometrial sampling and noticed that there was a 400cc fluid loss early in the case.The physician advanced a little further with good visualization but at some point the deficit jumped to 800cc.Upon procedure closure, the physician suspected perforation and a leak into the abdomen.A laparoscopy was done and a hole was found in the uterus with a little bit of fluid in the cavity.The physician stitched it up and suctioned any fluid out from the patient.Reportedly, the physician was unsure if the perforation was caused by the symphion resecting device or the dilation at the beginning of the case.The procedure was completed using this device.There were no patient complications as a result of the event.The patient condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHION¿ TISSUE REMOVAL SYSTEM RESECTING DEVICE
Type of Device
INSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7621710
MDR Text Key111762745
Report Number3005099803-2018-02003
Device Sequence Number1
Product Code PGT
UDI-Device Identifier08714729863656
UDI-Public08714729863656
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/05/2018
Device Model NumberFG-0201
Device Lot Number0065074600
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
Patient Weight66
-
-