One equator device was received in good condition with the power cord and hose.The filter was noted to be clean.The unit was powered on and tested at 36, 40, and 44 degree settings and was found to be operating as intended.An electrical safety and installed temperature test was also performed and measurements were found to be within specification.Upon further inspection, no evidence of arcing or burning was detected.Additionally, the device history record review was completed and no manufacturing issues were found.Based on the evidence and testing, the reported complaint allegation was not confirmed.
|