Model Number 8700-0650 |
Device Problems
Device Issue (2379); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/24/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll has not yet received the product for evaluation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
|
|
Event Description
|
On (b)(6) 2018, azm (b)(6) sale representative reported that during console demonstration, the thermogard xp ivtm system displayed a mid:14 (bg power supply) message after running for approximately 30 minutes.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint of mid:14 (bg power supply) error message was not confirmed during function test and during review of event log file on the thermogard xp ivtm system (sn (b)(4)).No physical damage observed on the console during visual inspection.The console passed all functional tests without any issue and it is ready for therapy use.Historical complaints were reviewed for service information related to the reported complaint and there were no similar complaints reported for thermogard console serial number (b)(4).
|
|
Search Alerts/Recalls
|