• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION ZOLL IVTM THERMOGARD XP; THERMAL REGULATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION ZOLL IVTM THERMOGARD XP; THERMAL REGULATING SYSTEM Back to Search Results
Model Number 8700-0650
Device Problems Device Issue (2379); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
Zoll has not yet received the product for evaluation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
 
Event Description
On (b)(6) 2018, azm (b)(6) sale representative reported that during console demonstration, the thermogard xp ivtm system displayed a mid:14 (bg power supply) message after running for approximately 30 minutes.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint of mid:14 (bg power supply) error message was not confirmed during function test and during review of event log file on the thermogard xp ivtm system (sn (b)(4)).No physical damage observed on the console during visual inspection.The console passed all functional tests without any issue and it is ready for therapy use.Historical complaints were reviewed for service information related to the reported complaint and there were no similar complaints reported for thermogard console serial number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM THERMOGARD XP
Type of Device
THERMAL REGULATING SYSTEM
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7626202
MDR Text Key112057606
Report Number3010617000-2018-00651
Device Sequence Number1
Product Code NCX
UDI-Device Identifier00849111075022
UDI-Public00849111075022
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072234
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0650
Device Catalogue Number8700-0650
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-