MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 5.0 CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 5.0 CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number SHORT
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.A visual and functional assessment was performed on the device which found that the device passed all tests.The device is being returned for further investigation.Once the investigation has been completed, a supplemental medwatch will be submitted.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the bearings of the attachment device were damaged.It was further determined that the device identification was unreadable.It was further determined that the ball bearings had fallen apart and the triangle symbol, color ring , balls and cage were missing.It was observed that the device failed lock operation and cutter insertion.It was noted in the service order that the burr devices slipped out of the device.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was not confirmed.A visual and functional assessment was performed and the device passed all operational specifications.Therefore, an assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 CM SHORT ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key7628836
MDR Text Key111989616
Report Number1045834-2018-51158
Device Sequence Number1
Product Code HBC
UDI-Device Identifier845384008798
UDI-Public(01)845384008798(11)101025
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSHORT
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-