• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the s5 gas blender system.The incident occurred in (b)(6).The involved gas blender has been returned to livanova deutschland for investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that an s5 gas blender system displayed an error message and audibly alarmed multiple times during in-service.There was no patient involvement.
 
Manufacturer Narrative
The s5 gas blender system was returned to livanova (b)(4) for further investigation.During the evaluation a defective front panel could be determined as root cause of the reported issue.The front panel was replaced and a new calibration was performed to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7629961
MDR Text Key112568674
Report Number9611109-2018-01065
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-