Boston scientific received information that this system was exhibiting noise, oversensing, pacing inhibition, increased shocking impedance measurements and pacing impedance measurements greater than 3000 ohms on the right ventricular (rv) channel.The issues noted on the rv channel started shortly after the patient had an av node ablation, upgrade procedure and a revision procedure shortly following the upgrade to secure the left ventricular (lv) lead in the device header due to pacing impedance measurements greater than 2000 ohms.The physician could not confirm if the rv lead had been manipulated during the lv lead revision.The reported issues could not be reproduced with in office testing and troubleshooting.Therefore, a revision procedure was performed and the lead was disconnected and reconnected to the device with normal impedance measurements and no observed noise.No additional adverse patient effects were reported.
|