Model Number BI-700-00027-120
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.The detector position louver was replaced.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding an imaging device being used outside of procedure.It was reported that the louvre cover was broken when setting up the operating room.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
sarah olhoft
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7634642
MDR Text Key112210331
Report Number1723170-2018-02970
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/29/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-