It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there were vertical lines on the central control monitor (ccm).The monitor and touch screen features were still operational.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) confirmed the reported complaint.After replacing the single board computer (sbc) board with a lab use only (luo) board the ccm powered up properly with no vertical lines.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|