MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAS
Device Problems Fluid/Blood Leak (1250); Failure to Prime (1492)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call the insulin pump would not exit the preparing to prime loop and insulin squirted out if the tubing during manual prime.The customer¿s blood glucose was unknown at the time of the incident.During troubleshooting, the customer changed the infusion set and reservoir, but insulin continued to drip after letting go of the act button during prime.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
Device passed the displacement test, basic occlusion test, occlusion test, prime test, rewind test and excessive no delivery test.No rewind anomaly noted.Device primed properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7635481
MDR Text Key112339104
Report Number3004209178-2018-84881
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169513860
UDI-Public(01)00643169513860
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA4523LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age67 YR
Patient Weight205
-
-