Model Number 170056
Device Problems Break (1069); Detachment Of Device Component (1104); Human-Device Interface Problem (2949)
Patient Problems Tissue Damage (2104); Radiation Exposure, Unintended (3164); No Code Available (3191)
Event Date 05/29/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopy, the first handle used had issues with loading and unloading and was difficult to toggle.They replaced it with second handle wherein it has no issues in loading and unloading however the needle broke and fell into the cavity of the patient.A new operation and an x-ray was performed to locate, remove the broken needle and correct the issue.The hospital stay was extended, there was tissue damage reported, and incision time was extended.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.No abnormalities were found during visual inspection and functional testing of the returned product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly or component related failures.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Needle was broken at suture swage.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly or component related failures.Replication of the bent or broken needle may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key7637520
MDR Text Key112305039
Report Number9612501-2018-01190
Device Sequence Number1
Product Code OCW
UDI-Device Identifier10884521126572
UDI-Public10884521126572
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model Number170056
Device Catalogue Number170056
Device Lot NumberJ7J2589X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age49 YR
-
-