NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number LEAD1058-50B
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2018
Event Type  Injury  
Manufacturer Narrative
The device was explanted but not returned.The manufacturing records were reviewed and no non-conformities were found.
 
Event Description
It was reported to nevro that during the trial procedure, one of the electrode separated from the lead into the subcutaneous tissue.No injuries were sustained by the patient.The lead was removed and the patient started the trial with one lead.The physician believed the electrode separation was due to the patient's anatomy.Since the patient had a successful trial, the physician has decided to remove the electrode during the permanent implant procedure.There have been no reports of further complications regarding this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Type of Device
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
catherine aronson
1800 bridge parkway
redwood city, CA 94065
6504333218
MDR Report Key7637929
MDR Text Key112333293
Report Number3008514029-2018-00275
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00813426020022
UDI-Public00813426020022
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date02/27/2021
Device Model NumberLEAD1058-50B
Device Catalogue NumberLEAD1058-50B
Device Lot Number94421062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age48 YR
-
-