|
Model Number PCOPM20 |
Device Problem
Torn Material (3024)
|
Patient Problems
Tissue Damage (2104); Obstruction/Occlusion (2422); No Code Available (3191)
|
Event Date 08/03/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, following a laparoscopic repair of a parastomal hernia procedure using a modified sugarbaker operative approach, the mesh was damaged.The surgeon observed a fracture of the mesh corresponding to the translucent stripe, where this originally covered the hernia defect.There was tissue damage and the patient experienced bowel obstruction as a result of the product problem.The surgical time was extended 30 minutes or more and the event also resulted to an extended patient hospital stay.The patient underwent reoperation for parastomal hernia to resolve the issue and to prevent permanent impairment.
|
|
Manufacturer Narrative
|
Evaluation summary: the visual examination of the provided picture shows that, a hole is visible in the middle of the mesh.Tackers are placed well, the distance from the edge of the mesh was found conform.The bowel is passing through the middle of the mesh by the hole.As no lot number was provided, a review of the device history record could not be performed.All process and test criteria are verified as complying with quality assurance (qa) specifications for all product lots prior release to market.The root cause could not be determined.The report has been added to our product complaints database which is monitored for similar occurrences.Based on our investigation and a complaint history review, the manufacture of the device is not suspected.There is no indication that there is a defective lot or that this event represents an emerging adverse quality trend.No immediate action required.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|