Model Number 8700-0650
Device Problems Device Issue (2379); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/19/2018
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint of the thermogard console did not power on was not confirmed during the initial functional testing.A visual inspection was performed and no discrepancies were observed.During functional testing, the reported complaint could not be reproduced, the thermogard console was able to power on with an old fuse.The system passed functional testing and it verified that the system met all the manufacturing specifications.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for thermogard with serial number (b)(4).
 
Event Description
As reported, during a self-test, the thermogard console (sn (b)(4)) was unable to power on.No patient involvement on this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM THERMOGARD XP
Type of Device
THERMAL REGULATING SYSTEM
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key7639642
MDR Text Key112441813
Report Number3010617000-2018-00670
Device Sequence Number1
Product Code NCX
UDI-Device Identifier00849111075022
UDI-Public00849111075022
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072234
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0650
Device Catalogue Number8700-0650
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-