The patient was implanted with a left ventricular assist device (lvad) on (b)(6) 2010.It was reported that the patient presented to clinic today due to the lvad system producing low flow alarms with pump stoppages.The patient was asymptomatic.Log files were reviewed by the technical service representative and pump stoppage events were confirmed.X-rays provided to the technical service representative revealed areas that were of concern, however they were not able to confirm that the area was responsible for the pump stoppage events.It was previously reported that on (b)(6) 2016, the patient received a prior distal end percutaneous lead (driveline) replacement, and there was not enough length of the driveline remaining to perform a second replacement.On (b)(6) 2018, the vad coordinator confirmed that the patient was having a pump exchange due to suspect of internal percutaneous lead (driveline) damage.The pump will be returned for analysis.No additional information was provided.
|
Manufacturer's investigation conclusion: the report of pump stoppages was confirmed based on the evaluation of the returned device.The pump was returned with the percutaneous lead (driveline) cut approximately 1.75¿ from the pump housing and the severed portion was returned.Examination of the pump portion of the driveline found it to be unremarkable.The brown wire on the severed portion of the driveline, approximately 2¿ from the pump housing, was damaged and broke off before continuity testing could be performed.Electrical continuity testing revealed discontinuities in the black, brown, red, yellow, and green wires.Visual inspection of the wires found insulation damage and wear in the yellow wire 2.5¿ from the pump housing and in the yellow, brown, and green wires 5.5¿ away from the pump housing.If the exposed conductors of the brown wire and either the yellow, or green wires made direct contact or simultaneous contact with the braided shield, the resulting phase-to-phase short could have caused the pump stop events captured on the submitted log files, while the system was operating on batteries.The pump stops that reportedly occurred while the system was operating on the power module could have occurred if any of the exposed conductors contacted the braided shield.The observed wire damage appeared to be the result of abrasion against the braided shield due to repetitive flexing.A review of the device history record revealed the device met applicable specifications.No further information was provided.The manufacturer is closing its file on this event.
|