• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN IDRIVE ULTRA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN IDRIVE ULTRA; STAPLE, IMPLANTABLE Back to Search Results
Model Number IDRVULTRA1
Device Problems Unintended Head Motion (1284); Misfire (2532); Difficult to Open or Close (2921)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a video-assisted thoracic surgery lobectomy, while the device was being applied to the pulmonary vein, when the device was fired, it stopped at the central part.After that, its adapter rotated automatically without any manipulation.The jaws of the device locked on tissue.The product was changed to complete the case.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one (1) device.Visual and functional evaluation of the handle noted no abnormalities.Evaluation of the eeprom noted that the code drive_motor_excessive_load_mode was present.This was indicative of a thick tissue application and could confirm the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the drive_motor_excessive _load_mode error codes may occur of the powered handle stops firing before the lower clamp reaches the distal end, which is likely due to the firing force reaching the upper design limit of the reload.This may occur under the following conditions.Application over tissue that is beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In either scenario, staples may not form properly and tissue may not be fully transected.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDRIVE ULTRA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7641659
MDR Text Key112457509
Report Number1219930-2018-03438
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884521142794
UDI-Public10884521142794
Combination Product (y/n)N
PMA/PMN Number
K121510
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIDRVULTRA1
Device Catalogue NumberIDRVULTRA1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Date Manufacturer Received03/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-