• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM Back to Search Results
Catalog Number 12673-05
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/11/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that an arteriotomy closure of the calcified left common femoral artery was attempted using a proglide device with an 8fr sheath after a interventional procedure (carotid stent).Reportedly, when the suture was harvested, a suture break was found.Another proglide device was used to achieve hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for analysis.The reported suture break was not confirmed as all components, including the suture, were not returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported difficulties and subsequent treatment appear to be related to an interaction of the device with patient anatomy or inability to maintain position/stability of the device during deployment due to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Type of Device
SUTURE MEDIATED CLOSURE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7642586
MDR Text Key112492265
Report Number2024168-2018-05081
Device Sequence Number1
Product Code MGB
Combination Product (y/n)N
PMA/PMN Number
P960043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Catalogue Number12673-05
Device Lot Number8011141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Date Manufacturer Received07/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SHEATH: 8FRHEPARIN; SHEATH: 8FRHEPARIN
Patient Outcome(s) Required Intervention;
-
-