Catalog Number 12673-05 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/11/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
It was reported that an arteriotomy closure of the calcified left common femoral artery was attempted using a proglide device with an 8fr sheath after a interventional procedure (carotid stent).Reportedly, when the suture was harvested, a suture break was found.Another proglide device was used to achieve hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).Evaluation summary: the device was returned for analysis.The reported suture break was not confirmed as all components, including the suture, were not returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported difficulties and subsequent treatment appear to be related to an interaction of the device with patient anatomy or inability to maintain position/stability of the device during deployment due to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|