• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
The primary report of a user advisory (ua) 12 (lifeband not present) was not reproduced during functional testing of the autopulse platform ((b)(4)); however, was confirmed through archive data review.The secondary report of the platform's drive shaft will not rotate was reproduced during functional testing.The root cause was a defective drive train.The archive data was reviewed and contained multiple ua 12 (lifeband not present) and ua 45 (driveshaft not at "home" position after power-on/restart) error messages.These error messages occurred due to the defective drive train.As part of routine service during testing, the platform was further examined a bent battery lock and damaged front enclosure.This physical damage found during visual inspection is a characteristic of mishandling of the device.This observation is not related to the reported event.After replacing the drive train, battery lock and front enclosure; the platform was further tested and passed all functional testing criteria and met all required specifications.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for the autopulse platform with serial number (b)(4).
 
Event Description
During shift check, the autopulse platform ((b)(4)) displayed a user advisory (ua) 12 (lifeband not present).Additionally, the platform's driveshaft will not rotate on one side.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7644297
MDR Text Key112574093
Report Number3010617000-2018-00673
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-