HILL-ROM BATESVILLE VERSACARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P3200K000679
Device Problems Device Alarm System (1012); Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/14/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the brake switch needed to be replaced.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in july 2017.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the brake switch to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the bed's brake not set alarm was not working.The bed was located on the 4th floor at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key7645100
MDR Text Key112870938
Report Number1824206-2018-00253
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Service Personnel
Type of Report Initial
Report Date 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3200K000679
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-