The hill-rom technician found the brake switch needed to be replaced.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in july 2017.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the brake switch to resolve the issue.Based on this information, no further action is required.
|