• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 37714
Device Problems Energy Output Problem (1431); Device Operates Differently Than Expected (2913); Insufficient Information (3190); Patient Device Interaction Problem (4001)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 01/01/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional via a manufacture representative regarding a patient implanted with an implantable neurostimulator (ins).It was reported that the patient's stim gets stronger sometimes when they are walking and that they have to stop and wait for it to 'slow down.' the rep reviewed with the patient that this was likely due to their adaptive stim setting for upright/mobile being set too high.The patient noted that they usually do not carry around their remote with them so they are unable to turn it down.The importance of having their remote with them in case this happens was reviewed with the patient.The patient was to meet with the rep on (b)(6) 2018.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was reported that it was confirmed that the cause of the stronger stimulation while walking around was due to the adaptive stimulation (as) settings.The mobile setting was adjusted and explained to the patient.The issue was resolved.No further information was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORE SENSOR
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7645837
MDR Text Key112600370
Report Number3004209178-2018-14547
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00613994610430
UDI-Public00613994610430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2013
Device Model Number37714
Device Catalogue Number37714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Date Device Manufactured12/15/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age43 YR
Patient Weight84
-
-